From Particle Formation...
... To Particle Characterisation

PoreMaster Series Porosimeters

Automatic Pore Size Analyzers


The PoreMaster consists of three automatic mercury intrusion porosimeters offering new concepts in automated pore size analysis.

The PoreMaster 33 porosimeter achieves a maximum pressure of 33,000 psia for pore size measurements in the range from over 950 micron to 0.0064 micron pore diameter. Two low pressure stations plus one high pressure station.

The PoreMaster 60 porosimeter generates pressure to 60,000 psia for pore size analysis from over 950 micron to 0.0036 micron pore diameter. Two low pressure stations plus one high pressure station. For greater sample throughput, you might be interested in the two high pressure stations offered by the PoreMaster GT Series.

The PoreMaster Macro porosimeter operates from approximately 0.2 psia to 50 psia for high-resolution pore size measurements in the range from over 950 micron to 4 micron pore diameter. Two low pressure stations only.

Functional Capability -60 & -60 GT -33 & -33 GT
PRESSURE RANGE    
0-50 psia >950-4.26 µm dia. >950-4.26 µm dia.
†Accuracy +/- 0.11% fso +/- 0.11% fso
Resolution 0.000763 psia 0.000763 psia
0-1500 psia 10.66-0.142 µm dia. 10.66-0.142 µm dia.*
†Accuracy +/- 0.11% fso +/- 0.11% fso
Resolution 0.0229 psia 0.0229 psia
0-Max. psia 10.66-0.0036 µm dia. 10.66-0.0064 µm dia.
†Non-Linearity +/- 0.05 % fso +/- 0.05 % fso
†Hysteresis +/- 0.10 % fso +/- 0.10 % fso
Resolution 0.916 psia 0.511 psia
PHYSICAL    
Depth 64.77 cm (25.50 in.) 64.77 cm (25.50 in.)
Width 51.44 cm (20.25 in.) 51.44 cm (20.25 in.)
Height 78.11 cm (30.75 in.) 91.44 cm (36 in.)
124.46 cm (lift extended)
Weight 180 kg (398 lbs.) 200 kg (440 lbs.)
Electrical 100-240V AC 50/60Hz 1000 VA

*optional
†Manufacturer’s Specifications

Read more about pore size distribution and analysis.

For more information contact us.

PoreMaster Porosimeters - Automatic Pore Size Analyzers

Poremaster Series

For more information contact us.